beautypg.com

LaMotte 2020 ClearTrace User Manual

Page 24

background image

Press ENTER to initiate zero
calibration. A zero calibration
may take some time, as the
zero must be calibrated over a
wide dynamic range. During
this process the calibrating...
message is shown.

When the zero calibration is
complete the user will be
asked to confirm the reult
with the ENTER key.

After pressing ENTER to
confirm Calibration is okay,
the Measure screen will be
displayed.
NOTE: 0.000 NTU may not
be displayed. An acceptable
reading could be as high as
0.015 NTU.

If an error occurs during
calibration, the reason for the
failure is shown.

Press ENTER to acknowledge
the calibration failure and
return to the measure screen.

23

0.045 NTU

AB

13.1 °C

DS1:

6

MENU

ESC

ENTER

t

t

t

t

Zero
Calibrating…

MENU

ESC

ENTER

t

t

t

t

Zero
Cal Fail

MENU

ESC

ENTER

t

t

t

t

Zero
Confirm cal ok?

MENU

ESC

ENTER

t

t

t

t

0.000 NTU

AB

13.1 °C

DS1:

6

MENU

ESC

ENTER

t

t

t

t