6. CALIBRATION

NOTE: The unit should be on for 30
minutes to allow the lamp to warm up
before begining calibrations or taking
measurements.

6.1. Span Calibration

To calibrate the span for the Turbidity
System:

1. Drain the sample chamber.
2. Shut off flow to the sample chamber.
3. Remove the chamber lid and place it

on top of the AV38.

4. Rinse the chamber with clean water and then with a small amount of

calibration solution. (See calibration kit instructions Code-5-0114 or
code 5-0015 for procedure.)

5. Shut off the Drain.

6. Fill the sample chamber with AMCO Clear Turbidity Standard, 20

NTU calibration solution. The standard should overflow the
OUTLET weir located about ¾ of the way to the top of the sample
chamber. As an alternative a formazin 20 NTU calibration standard
may be prepared from a stock solution of 4000 NTU formazin and
used within 24 hours. Refrigerate when stored.

7. After filling the sample chamber with the standard, wait for the

reading to settle (less than a minute depending on the filter setting).

8. Perform a 1-point

sample calibration as
shown in the following
menu tree.

The complete calibrate menu
for the turbidity and
suspended solids is shown.

19

Drain Chamber, Shut off Inlet

Drain and Inlet are shut off:

Ready to calibrate

Sample Chamber

with Lid Removed

Be careful to pour standard directly

into the SAMPLE CHAMBER.

Do not pour standard into the

Bubble trap.

During this operation, water will

remain int he BUBBLE TRAP and

it is not necessary to remove it.

Sample Chamber

Bubble Trap

Calibrate
1-Point Sample

MENU

ESC

ENTER

Zero
Temperature
Exit

t

t

t

t

t

t

t

t

Figure 5.15: The complete calibrate menu for turbidity