If the reading displays

Er2

, repeat procedure on diluted sample, and multiply the result

by the appropriate dilution factor. See 1200 Colorimeter Instruction Manual for
procedure.

10.

9.

Cap tube and invert

to mix.

11.

Align the index line with the

arrow on the meter, insert

tube into chamber.

Close the lid. Push

the

button. Record

results as ppm

Manganese.

READ

Note:

For the best possible results, carry a

reagent blank through the

procedure. After scanning the blank

in Step 4, perform the test procedure

on clear, colorless, distilled or

deionized water. Subtract results of

reagent blank from all subsequent

test results.

12.

Use the 0.5 mL

pipet (0369) to

add 0.5 mL of

*Manganese

Indicator Reagent

(3956).

Wipe tube dry.

CAL

I200

COLORIMETER

••••••••••••••••••

READ

L Mott