OZONE TEST PROCEDURE - INDIGO METHOD

Read the 1200 Colorimeter Manual before proceeding.
Carefully wipe tubes dry before inserting into the
colorimeter chamber.

PROCEDURE

2

Use the 1.0 mL

transfer pipet (2-

2170) and pipette

pump (2-2216) to

add 1.0 mL of HR

Reagent to each of

2 clean colorimeter

tubes (0967).

If chlorine is

present, add 3

drops Chlorine

Inhibitor (3990)

to each tube.

Cap tubes.

8.

5.

3.

4.

Use the 50 mL graduated

cylinder (0418) to carefully

add 45 mL of *Ozone Buffer

(3991) to amber glass bottle

marked HR Reagent (0680).

Use the 5 mL

volumetric pipet

(0329) and pipette

pump (2-2216) to add

5 mL of Indigo Blue

Stock Solution (3989)

to the amber glass

bottle. Cap and mix.

1.

2.

7.

Take one of the

prepared colorimeter

tubes and sampling

apparatus (0681) to

sampling site.

6.

Lower the end of the tubing

of sampling apparatus to

desired depth. Slowly

withdraw and depress

plunger several times to

purge syringe and tubing.

Slowly withdraw plunger to

fill purged syringe.

Remove plastic tubing

from syringe. Remove

Fill the second prepared

tube (0967) to the 10

mL line with ozone free

water. This is the

Reagent Blank.

HR Reagent

cap from the prepared tube.

Place tip of syringe against

inside of the prepared tube.

Slowly depress plunger and

fill to the 10 mL line and

cap. This is the Sample

Tube. NOTE: Do not shake

or invert the sample.