OZONE TEST METHOD SPECIFICATIONS

APPLICATION

Drinking, pool, and aquatic waters.

RANGE

0.0 - 0.5 ppm Ozone

METHOD

Ozone rapidly and stoichiometrically decolorizes Indigo Trisulfonate under acidic
conditions.

HANDLING & PRESERVATION

Ozone is extremely unstable in aqueous solutions. Test must be performed
immediately and the sample must not be agitated.

INTERFERENCES

Manganese at any level will interfere.