beautypg.com

LaMotte 1200 Colorimeter User Manual

Page 31

background image

er2

TOTAL RESIDUAL CHLORINE

If either chlorine reading displays

, repeat procedure on diluted sample, and

multiply the result by the appropriate dilution factor. See 1200 Colorimeter Instruction
Manual for procedure.

Levels of chlorine above 10 mg/L will cause a bleaching effect on the DPD indicator, and
may give a false indication that no chlorine is present. If it is possible that the chlorine
concentration is greater than 10 mg/L (e.g. after shock treatment), perform test on a
diluted sample and multiply the result by appropriate dilution factor.

CAUTION: DO NOT leave reacted DPD samples in test tubes. Discard sample and
thoroughly rinse tubes. If allowed to remain, DPD will stain tubes, significantly
impairing the operation of the 1200 Colorimeter. If necessary, acid wash, and vigorously
clean glassware with test tube brush and detergent.

**For wastewater samples, Standard Methods for the Examination of Water and Wastewater
recommends waiting 2 minutes for full color development.

11.

9.

10.

12.

••••••••••••••••••

READ

2.55

ZERO

COLORIMETER

I200

Cap and mix

until tablet

disintegrates.**

Push the

READ

button.

Record reading

as ppm Total

Residual

Chlorine.

Insert the tube

into the

chamber, being

sure to align the

index line with

the arrow on

the meter.

Close the lid.

Subtract the Free Available

Chlorine reading from the

Total Residual Chlorine

reading to determine ppm

Combined Chlorine

(Monochloramine,

Dichloramine, and Nitrogen

Trichloride) present in the

water sample.