COPPER TEST PROCEDURE - DIETHYLDITHIOCARBAMATE

Read the 1200 Colorimeter Manual before proceeding. Carefully
wipe tubes dry before inserting into the colorimeter chamber.

COPPER

Fill the Water
Sample Collecting

Bottle (0688) with

sample water. This

will be used to
dispense sample

water for the tests.

Insert the tube
into the

chamber, being

sure to align the

index line with
the arrow on

the meter.

Close the lid.

This tube is the

blank or zero.

Rinse and fill a
colorimeter tube

(0290) to the 10

mL line with

sample water.
Cap and wipe

dry.

5.

3.

1.

2.

4.

6.

I200

COLORIMETER

••••••••••••••••••

READ

bla

ZERO

7.

Align the index line with the
arrow on the meter, insert tube

CAL

I200

COLORIMETER

••••••••••••••••••

READ

into chamber.
Close the lid. Push

the
button. Record

results as ppm

Copper.

READ

Cap and invert to
mix. Wipe tube dry.

Remove tube from
colorimeter. Add 5

drops of *Copper 1

(6446).

Push the

button

to turn the meter

on. Press the

button

and hold it for 2

seconds until

is displayed.

Release the button

to take a blank

reading (0.0 ppm).

READ

ZERO

bla