Calibration procedure – LaMotte 1200 Colorimeter User Manual
Page 18
![background image](/manuals/596663/18/background.png)
CALIBRATION PROCEDURE
The 1200 has been pre-calibrated. Recalibration of the 1200 by the user is not
required. However, a procedure to standardize the calibration (shown below)
should be performed to obtain the most accurate readings.
18
Prepare standard
solutions to be
tested.
Insert the tube
into the
chamber, being
sure to align the
index line with
the arrow on
the meter.
Close the lid.
This tube is the
blank or zero.
Rinse and fill a
colorimeter tube
(0290) to the 10
mL line with
standard
solution. Cap
and wipe dry.
7.
8.
5.
3.
1.
2.
4.
6.
React a standard
that is in the range
of the reagent
system, and similar
to the expected
range of the sample.
Follow the
individual test
procedure.
Wipe the tube clean with a
lint-free cloth.
2.50
ppm
Push the
button to turn the
meter on. Press the
button
and hold it for 2
seconds until
is displayed.
Release the button
to take a zero
reading (0 ppm).
Insert the
tube into the
chamber,
being sure to
align the
index line
with the
arrow on the
meter. Close
the lid.
Push the
button. If the
displayed value is not the same as
the value of the
reacted standard
(within the
specification
limits), continue
with the
calibration
procedure.
I200
••••••••••••••••••
ZERO
COLORIMETER
READ
2.55
I200
••••••••••••••••••
ZERO
COLORIMETER
READ
bla
READ
READ
ZERO