CHLORINE TEST PROCEDURE - DPD METHOD

Read the 1200 Colorimeter Manual before proceeding. Carefully
wipe tubes dry before inserting into the colorimeter chamber.

FREE AVAILABLE CHLORINE

Insert the tube

into the

chamber, being

sure to align the

index line with

the arrow on

the meter.

Close the lid.

This tube is the

blank or zero.

8.

5.

3.

4.

Fill the Water

Sample Collecting

Bottle (0688) with

sample water. This

will be used to

dispense sample

water for the tests.

Rinse and fill a

colorimeter tube

(0290) to the 10

mL line with

sample water.

Cap and wipe

dry.

1.

2.

I200

COLORIMETER

••••••••••••••••••

READ

bla

ZERO

7.

CAL

I200

COLORIMETER

••••••••••••••••••

READ

Align the index line

with the arrow on

the meter, insert

tube into chamber.

Close the lid. Push

the

button.

Record results as

ppm Free Available

Chlorine.

READ

Remove the tube and

add 5 drops DPD #1A

Free Chlorine Reagent

(6740), and 5 drops of

*DPD #1B Free

Chlorine Reagent

(6741).

6.

Cap and invert to

mix. Wipe tube dry.

Make readings

within 30 seconds.

Push the

button to turn the

meter on. Press the

button and

hold it for 2 seconds
until

is

displayed. Release

the button to take a

zero reading
(0.00 ppm).

READ

ZERO

bla

Save sample for the

Total Residual

Chlorine Test.

Proceed to Step 9.