PROCEDURE
1.
Rinse a clean colorimeter tube (0967) with sample water. Fill to the 10 mL line with sample.
2.
Select setting 2 on “Select Wavelength” knob and press “30 Second Read” button.
3.
Insert tube into colorimeter chamber and adjust to 100%T with “Set Blank” knob. This is the %100T blank.
4.
Remove tube from colorimeter. Add 6 drops *Silica Reagent #1 (V-4466). Cap and invert to mix.
5.
Add 12 drops of *Silica Reagent #2 (V-4467). Cap and mix. Wait 5 minutes.
6.
At end of 5 minute waiting period, add 2 drops of *Silica Reagent #3 (V-4468). Cap and mix.
7.
Press “30 Second Read” button and insert tube into colorimeter chamber. Record %T as soon as reading stablilizes.
8.
Consult calibration chart to determine silica concentration in parts per million (ppm).
DC1600 SILICA ~ HIGH RANGE CALIBRATION CHART
%T
9
8
7
6
5
4
3
2
1
0
90
1
2
3
4
5
7
8
9
10
11
80
13
14
15
17
18
19
20
22
23
25
70
26
27
29
30
32
33
35
36
38
40
60
41
43
44
46
48
50
51
53
55
57
50
59
61
63
65
67
69
71
73
75
40
30
20
10
0
NOTE: To extend range to 150 ppm, perform a 2:1 dilution of water sample with silica free water. Perform test and
multiply result by 2.