HYDRAZINE

p-DIMETHYLAMINOBENZALDEHYDE METHOD

CODE 3656

QUANTITY

CONTENTS

CODE

2 x 60 mL

*Hydrazine Reagent A

*4841-H

10 g

*Hydrazine Reagent B Powder

*4842-D

1

Pipet, 1.0 mL, plastic

0354

1

Spoon, 0.1 g, plastic

0699

*WARNING: Reagents marked with a * are considered hazardous substances. Material Safety Data Sheets (MSDS) are supplied for

these reagents. For your safety, read label and accompanying MSDS before using.

Hydrazine, N

2

H

4

, is added to the water in high pressure boilers to reduce corrosion of pipes and fittings by acting as an

oxygen scavenger.

APPLICATION:

Water and boiler water, industrial wastewater.

RANGE:

0 – 1.0 ppm Hydrazine

METHOD:

p-Dimethylaminobenzaldehyde reacts with hydrazine under acidic conditions to form a yellow

color in proportion to the amount of hydrazine present.

SAMPLE HANDLING

& PRESERVATION:

Samples should be analyzed as soon as possible after collection due to the ease with which

hydrazine becomes oxidized. Acidification of the sample may increase the time between

collection and analysis.

INTERFERENCES:

The substances normally present in water do not interfere with the test, with the exception of

strong oxidizing agents.

PROCEDURE

1.

Rinse a clean colorimeter tube (0967) with sample water. Fill to the 10 mL line with sample.

2.

Select setting 2 on “Select Wavelength” knob and press “30 Second Read” button.

3.

Insert tube into colorimeter chamber and adjust to 100%T with “Set Blank” knob. This tube is the 100%T blank.

4.

Remove tube from colorimeter. Use the 1 mL pipet (0354) to transfer 4 mL of *Hydrazine Reagent A (4841). Cap and

mix.

5.

Use the 0.1 g spoon (0699) to add one measure of *Hydrazine Reagent B Powder (4842). Cap and mix. Wait 10

minutes for maximum color development. An undissolved portion of Hydrazine Reagent B may remain in bottom of

tube without adversely affecting results.

6.

At end of 10 minute waiting period, press “30 Second Read” button and insert tube into colorimeter chamber. Record

%T as soon as reading stabilizes.

7.

Consult calibration chart to determine hydrazine concentration in parts per million (ppm).

DC1600 HYDRAZINE CALIBRATION CHART

%T

9

8

7

6

5

4

3

2

1

0

90

0.01

0.01

0.02

0.02

0.02

0.03

0.03

0.03

0.04

0.04

80

0.04

0.04

0.05

0.05

0.05

0.06

0.06

0.06

0.07

0.07

70

0.07

0.08

0.08

0.09

0.09

0.09

0.10

0.10

0.10

0.11

60

0.11

0.12

0.12

0.12

0.13

0.13

0.14

0.14

0.15

0.15

50

0.16

0.16

0.17

0.17

0.18

0.18

0.19

0.19

0.20

0.20

40

0.21

0.21

0.22

0.23

0.23

0.24

0.24

0.25

0.26

0.26

30

0.27

0.28

0.29

0.29

0.30

0.31

0.32

0.33

0.34

0.35

20

0.36

0.37

0.38

0.39

0.40

0.41

0.42

0.43

0.45

0.46

10

0.48

0.49

0.51

0.52

0.54

0.56

0.58

0.61

0.63

0.66

0

0.69

0.72

0.76

0.81

0.86

0.92

1.01

41