REV6 OPERATION / MAINTENANCE MANUAL

28

CONTENTS

5.1.4 Error Display on the Integrated Panel

Table 9 - Errors and Warnings with Indication on Standalone Controller Display

Error
Source

Errors

Error Code
on Display

Motor

No motor

Er 01

Motor

Motor cable (power wires) not connected to controller

Er 02

Motor

Motor cable (sensor wires) not connected to controller

Er 03

Motor

No impeller

Er 04

Controller Short circuit

Er 05

Controller Over current in the bearing coils

Er 06

Controller Over current in the drive coils

Er 07

Controller DC-Link voltage out of range (< 40 or > 54 V DC)

If the voltage is out of range the system starts to reduce the speed and a warning is generated. When
reaching 0 rpm and the voltage is still out of range the system is disabled and an error is generated. In
case the voltage is again within the range during speed reduction the system switches to normal
operation and no Error is generated.

Er 08

Controller Communication problems EEPROM Controller

Er 09

Motor

Communication problems EEPROM Motor

Er 10

Controller Controller temp. over 80°C or more than 10 minutes above 70°C

Er 11

Motor

Motor temp. over 100°C or more than 10 minutes above 90°C

Er 12

Pump

Dry running of pump circuit:

-> Pump keeps running on reduced speed (5000 rpm) -> The system accelerates to the original speed
value when the pump is refilled with liquid. -> Note that the speed is only reduced during dry running if
the pump speed was ≥ 6000 rpm.

Blinking

dots on

display

- In case of an error the system can only be restarted with a reset or a power supply restart

- Standard firmware is D6.25
- For other configurations of error codes refer to alternate controller or firmware documentation