8. This completes the offset configuration.
9. At this point, the instrument will continue through the configuration (CONFIG) mode
of the instrument or press MODE/EXIT to return to the AUTO mode.
6.1 Restoring Factory Settings
If the instrument is unable to perform a calibration due to a low lamp output, bad
calibration standard or a dirty cuvette, the instrument will display CAL on the lower row
of the display and Err on the upper row. The operator has two choices to correct this
problem. If the operator can determine whether a poor calibration or a low lamp caused
the problem, he/she can remedy the problem and recalibrate. If all else fails, the operator
may restore the factory calibration and configuration settings by performing the following
operation. Push and hold the
tbutton. Now push and release the ↵ then release the
tbutton. Factory calibration and factory configuration have now been restored.
Note: Restoring the factory settings allows the use of the AccUView with reduced accuracy.
The original problem still exists and must be determined and corrected before
accurate operation of the AccUView will be resumed.