0 instrument calibration – HF scientific AccUView UV %Transmission Monitor User Manual
Page 15
4.3 Security Access Feature
The instrument is equipped with a security access code feature that can be activated in the
configuration mode. If the security feature is enabled, the screen shown in the illustration
below will appear when the MODE/EXIT button is pressed.
The security code (333) must be entered to gain access to CAL or CONFIG menus.
Notice that the first number in the code is flashing. The flashing indicates that this is the
number to be entered. Use the
t
or
u arrows to select the first of the three numbers in
the code and then press the
↵ button to accept the first number of the code. Now enter the
second number in the code. Proceed as with the first number followed by
↵. Then repeat
the process for the third number in the access code, and finish with the
↵
button.
If the valid access code has been selected, the instrument will be directed to the calibration
mode. If the CONFIG mode is needed push the MODE/EXIT button one more time.
If the wrong access code is selected, the instrument will return to the AUTO mode. Refer
to section 7.6 Enabling the Security Access for more information.
AccUView (6/10)
Rev. 4.0
10