6 mounting the capillary, Mounting the capillary – Metrohm Multi-Mode Electrode pro User Manual
Page 43
■■■■■■■■■■■■■■■■■■■■■■
5 Maintenance
6.1246.120 Multi-Mode-Electrode pro
■■■■■■■■
35
5.2.6
Mounting the capillary
Caution
In this procedure mercury is used. Always carry out the following steps
in a dry drip pan.
There are two types of capillaries which can be mounted into the MME
pro, depending on the application:
■
Silanized capillaries (6.1226.050) are especially suitable for voltammet-
ric determinations at the HMDE in acidic to slightly alkaline solutions
which do not contain organic solvents.
■
Non-silanized capillaries (6.1226.030) are especially suitable for polaro-
graphic determinations at the DME or SMDE and for voltammetric
determinations at the HMDE in all types of solutions, especially alkaline
solutions with pH>10 or in solutions containing solvents, where silan-
ized capillaries cannot be used.
Mounting the capillary
To carry out this procedure you need the MME pro and the following
accessories:
■
Drip pan (6.2711.030)
■
Electrode stand (6.2615.130)
■
Wrench (6.1247.220)
■
Silanized capillary (6.1226.050) or non-silanized capillary (6.1226.030)
■
Seal (6.1247.150)
■
Scissors
Caution
In this procedure mercury is used. Always carry out the following steps
in a dry drip pan.