3 Preparing the MME pro for use
6.1246.120 Multi-Mode-Electrode pro
■■■■■■■■
17
Note
If you use another software than 797 VA Computrace, then follow the
instructions on adjusting the needle valve in the manual of your instru-
ment.
For adjusting the needle valve you need the wrench (6.1247.220) or a
coin which will fit the slot of the screw.
1 Preparation
■
Place an empty measuring vessel.
■
Set the inert gas supply to 1.0 to 1.2 bar, in order to pressurize
the electrode.
■
In the 797 VA Computrace software, open the dialog Compu-
trace Control with Utility
▶ Computrace Control and select
the electrode type HMDE.
■
As soon as a drop of mercury appears at the tip of the capillary,
fill approximately 20 mL of water and a few drops of KCl-solution
into the measuring vessel.
2 Rough adjustment
■
With the wrench, rotate the screw of the needle holder (1-1)
clockwise until the mercury stops dropping.
Check: After each half rotation of the screw, knock off the drop
and check if a new drop is forming.
3 Fine adjustment
■
Rotate the screw slowly counterclockwise until a new drop is
forming.
■
Reseal the capillary with a small rotation in clockwise direction
(ca. 10°).
■
Rotate the screw clockwise by another quarter rotation (90 –
100°).
The MME pro is now ready for operation.