3 removing the capillary, Removing the capillary – Metrohm Multi-Mode Electrode pro User Manual
Page 39

■■■■■■■■■■■■■■■■■■■■■■
5 Maintenance
6.1246.120 Multi-Mode-Electrode pro
■■■■■■■■
31
■
Drip pan (6.2711.030)
■
Electrode stand (6.2615.130)
■
Wrench (6.1247.220)
■
Mercury collection container
■
Mercury drop catcher (6.2406.000)
1 Placing the MME pro into the electrode stand
■
Place the electrode stand into the drip pan.
■
Place the MME pro into the electrode stand.
The mercury refill opening is pointing upwards (see Figure (2-B),
Page 7).
2 Opening the mercury reservoir
■
Use the wrench to remove the screw cap from the mercury refill
opening.
■
Use the mercury drop catcher to collect mercury adhering to the
screw cap.
■
Place the screw cap into the drip pan.
3 Pouring out the mercury
■
Pour the mercury from the mercury reservoir into the mercury col-
lection container.
■
Knock the MME pro lightly to remove remaining mercury from
the hollow spaces and pour it into the mercury collection con-
tainer.
Collect small amounts of dispersed mercury with the mercury drop
catcher.
5.2.3
Removing the capillary
Caution
In this procedure mercury is used. Always carry out the following steps
in a dry drip pan.
To remove the capillary you need the following accessories:
■
Drip pan (6.2711.030)
■
Electrode stand (6.2615.130)
■
Mercury drop catcher (6.2406.000)