3 Preparing the MME pro for use
6.1246.120 Multi-Mode-Electrode pro
■■■■■■■■
15
■
Mercury
■
Mercury drop catcher (6.2406.000)
1
2
3
4
Figure 5
MME pro – Filling in mercury
1 Placing the MME pro into the electrode stand
■
Place the electrode stand into the drip pan.
■
Place the MME pro into the electrode stand.
The mercury refill opening is pointing upwards.
2 Opening the screw cap
■
Open the screw cap (1-7) at the mercury refill opening with the
wrench and remove it.
3 Filling in mercury
Using the syringe, carefully fill in the mercury into the mercury reser-
voir.
Note
Only fill in mercury until it reaches the making max. fill.