6 Troubleshooting

872 Extension Module Suppression

■■■■■■■■

41

Problem

Cause

Remedy

SPM – reduced capacity.

Regenerate the SPM (see Chapter 5.4.4.2,
page 33)
.

SPM – No (or insuffi-
cient) pumping of
regeneration or rins-
ing solution

Leak in the system.

Check connections.

Peristaltic pump – contact
pressure too weak.

Correctly set contact pressure (see "Set the
contact pressure", page 19)
.

Peristaltic pump – filter
blocked .

Replace the filter (see "Replacing the filter",
page 30)
.

SPM – backpressure too
high.

Clean the SPM (see Chapter 5.4.4.3, page 35)
or replace parts (see Chapter 5.4.4.4, page
36)
.

Peristaltic pump – pump
tubing defective.

Replace pump tubing.