6 troubleshooting, 1 problems and their solutions, Problems and their solutions – Metrohm 872 Extension Module Suppression User Manual
Page 48
6.1 Problems and their solutions
■■■■■■■■■■■■■■■■■■■■■■
40
■■■■■■■■
872 Extension Module Suppression
6 Troubleshooting
6.1
Problems and their solutions
Problem
Cause
Remedy
Marked rise in pres-
sure
Suppressor – blocked.
■
Regenerate the suppressor (see Chapter
5.4.4.2, page 33).
Note: 6.2821.180 pump tubing connection
with filter must be used.
Peristaltic pump –
insufficient or no
delivery rate
Peristaltic pump – contact
pressure too weak.
Correctly set contact pressure (see "Set flow
rate", page 19).
Peristaltic pump – filter
blocked.
Replace the filter (see Chapter 5.3.2.2, page
30).
Peristaltic pump – pump
tubing defective.
Replace pump tubing (see Chapter 5.3.2.1,
page 29).
Background conduc-
tivity too high
Suppressor – not connec-
ted.
Connect the suppressor (see Chapter 3.5,
page 19).
Suppressor – regeneration
or rinsing solution flow
problems.
Check the flow of the regeneration and rinsing
solution (see Chapter 3.5.2, page 23).
Suppressor – No (or
insufficient) pump-
ing of regeneration
or rinsing solution
Leak in the system.
Check connections.
Peristaltic pump – contact
pressure too weak.
Correctly set contact pressure (see "Set the
contact pressure", page 19).
Peristaltic pump – filter
blocked .
Replace the filter (see "Replacing the filter",
page 30).
Suppressor – backpressure
too high.
Clean the suppressor (see Chapter 5.4.4.3,
page 35) or replace parts (see Chapter 5.4.4.4,
page 36).
Peristaltic pump – pump
tubing defective.
Replace pump tubing.
Greater rise in the
baseline
Suppressor – reduced
capacity.