6 troubleshooting, 1 problems and their solutions, Problems and their solutions – Metrohm 872 Extension Module Suppression User Manual

Page 48

background image

6.1 Problems and their solutions

■■■■■■■■■■■■■■■■■■■■■■

40

■■■■■■■■

872 Extension Module Suppression

6 Troubleshooting

6.1

Problems and their solutions

Problem

Cause

Remedy

Marked rise in pres-
sure

Suppressor – blocked.

Regenerate the suppressor (see Chapter
5.4.4.2, page 33)
.

Note: 6.2821.180 pump tubing connection
with filter must be used.

Peristaltic pump –
insufficient or no
delivery rate

Peristaltic pump – contact
pressure too weak.

Correctly set contact pressure (see "Set flow
rate", page 19)
.

Peristaltic pump – filter
blocked.

Replace the filter (see Chapter 5.3.2.2, page
30)
.

Peristaltic pump – pump
tubing defective.

Replace pump tubing (see Chapter 5.3.2.1,
page 29)
.

Background conduc-
tivity too high

Suppressor – not connec-
ted.

Connect the suppressor (see Chapter 3.5,
page 19)
.

Suppressor – regeneration
or rinsing solution flow
problems.

Check the flow of the regeneration and rinsing
solution (see Chapter 3.5.2, page 23).

Suppressor – No (or
insufficient) pump-
ing of regeneration
or rinsing solution

Leak in the system.

Check connections.

Peristaltic pump – contact
pressure too weak.

Correctly set contact pressure (see "Set the
contact pressure", page 19)
.

Peristaltic pump – filter
blocked .

Replace the filter (see "Replacing the filter",
page 30)
.

Suppressor – backpressure
too high.

Clean the suppressor (see Chapter 5.4.4.3,
page 35)
or replace parts (see Chapter 5.4.4.4,
page 36)
.

Peristaltic pump – pump
tubing defective.

Replace pump tubing.

Greater rise in the
baseline

Suppressor – reduced
capacity.

Regenerate the suppressor (see Chapter
5.4.4.2, page 33)
.