5 Handling and maintenance

872 Extension Module Suppression

■■■■■■■■

37

Replacing parts of the suppressor

Replace the parts of the suppressor/SPM as follows:

1 Disconnecting the suppressor/SPM from the IC system

Switch off the instrument.

Disconnect all capillaries of the suppressor/SPM from the IC sys-
tem.

2 Dismantling the suppressor/SPM

Unscrew union nut (11-1) from the housing .

Pull the connecting piece (11-2) and the rotor (11-3) out of the
housing.
The connecting piece and the rotor normally stick to one another
– if this is not the case: Take a sharp object, inserti it into the slot
of the housing (11-5) and pull out the rotor in this way.

Detach the connecting piece from the rotor.

3 Cleaning the new rotor

Clean the surface of the rotor (11-3) with ethanol using a lint-free
cloth.

4 Inserting the new rotor

Caution

An incorrectly inserted rotor can be destroyed during start-up.

Insert the new rotor (11-3) into the housing (11-4) in such a way
that the tubing connections on the rear of rotor fit into the corre-
sponding recesses inside the housing and one of the three holes
of the rotor is visible from below in the slot of the housing (11-5).

If the rotor is correctly inserted, its sealing surface will be approx.
4 mm within the housing. If this is not the case, the rotor must be
moved into the right position from below using a sharp object
(e.g. screwdriver).

5 Cleaning the new connecting piece

Clean the surface of the rotor (11-2) with ethanol using a lint-free
cloth.