2 regenerating the suppressor, Figure 11, Components – Metrohm 872 Extension Module Suppression User Manual

Page 41

background image

■■■■■■■■■■■■■■■■■■■■■■

5 Handling and maintenance

872 Extension Module Suppression

■■■■■■■■

33

1

2

3

4

5

Figure 11

Components

1

Union nut

2

Connecting piece (6.2835.010)

3

Rotor

4

Housing

5

Slot in the housing

5.4.4.2

Regenerating the suppressor

If the suppressor units or the neutralization units of the SPM are loaded
for a longer period with certain heavy metals (e.g. iron) or organic impuri-
ties, these can no longer be completely removed with the standard regen-
eration solution. The capacity of the suppressor units is consequently
affected, which can result in reduced phosphate sensitivity in less serious
cases and a significant rise in the baseline in more serious cases. The neu-
tralization units of the SPM react to contamination with a reduced verifica-
tion sensitivity.

If such capacity or sensitivity problems occur at one or more positions, all
suppressor units or neutralization units must be regenerated wit one of
the following solutions::

Contamination with heavy metals:
1 mol/L H

2

SO

4

+ 0.1 mol/L oxalic acid

Contamination with organic cationic complexing agents:
0.1 mol/L H

2

SO

4

/ 0.1 mol/L oxalic acid / acetone 5%

Severe contamination with organic substances:
0.2 mol/L H

2

SO

4

/ acetone

≥ 20%

Caution

The pump tubings made of PVC must not be used for rinsing with solu-
tions containing organic solvents.

We recommend to use the high pressure pump for regeneration.