3 Installation

872 Extension Module Suppression

■■■■■■■■

17

3 Connecting the pressure side

Note

Depending on the use of the peristaltic pump, on the pressure side
you can either connect:

Case A: a 6.2744.180 pump tubing connection with filter
(see Figure 6, page 17) or

Case B: a 6.2744.160 pump tubing connection without filter
(see Figure 7, page 17).

For pumping the auxiliary solutions to the MSM or to the SPM, a
6.2744.180 pump tubing connection with filter must be used.

Case A: 6.2744.180 pump tubing connection with filter:

1

2

3

Figure 6

Install pump tubing connection with filter

1

Union nut

2

Adapter

3

Tubing olive with filter holder

Slide union nut (6-1) onto the pump tubing.

Select a suitable adapter (6-2) and slide it onto the pump tubing.
The type of adapter depends on the pump tubing (see Table 1,
page 18)
.

Place the tubing olive with filter holder (6-3) onto the pump tub-
ing.

Screw the union nut (6-1) onto the tubing olive (6-3).

or

Case B: 6.2744.160 pump tubing connection without filter:

1

2

3

Figure 7

Install pump tubing connection without filter