Metrohm 871 Advanced Bioscan User Manual
Page 29

3.4 Optimization of the measuring potential
871 Advanced Bioscan / Instructions for Use 8.871.1003
21
The final aim is to select a suitable potential for use as the measuring
potential for pulsed amperometric detection. This means that usually a
compromise has to be found between the highest possible sensitivity,
selectivity and reproducibility:
Higher potentials can increase the sensitivity; however, more sub-
stances can also be determined and this reduces the selectivity. In or-
der to achieve high reproducibility, a potential from a flatter segment
near the maximum of the analyte voltammogram should be selected. In
the example shown in Figure 6, E
1
would be such a potential deter-
mined in this way.
Hydrodynamic
Scan
Situation
•
Selectivity for an inadequate
•
Sensitivity of a substance
voltammogram
voltammogram
IC-separation is to be im-
proved
peak is to be increased
Requirement
•
pure form
Advantages
tions, i.e. a direct check of
•
Several substances can be
substances
Disadvantages
de
Substance can be investi-
gated in the mixture
•
Substance must be present
in a
•
IC parameters must be
known
•
"Chromatographic" condi-
•
Less time needed for single
selectivity and sensitivity is
possible
investigated at the same
time
•
More time required for a
single substance
•
Possible formation of an ox-
ide layer on the working
electro
During the determ
ination of the optimum measuring potential the con-
stancy of the parameters pH and temperature as well as the pulsation-
free eluent supply are important (see Section 4.2).