2 Installation

2.2.1 Fuses

Fuses 2.5AT/250V are installed in fuse holder 7 of the 871 Advanced

Bioscan.

871 Advanced Bioscan / Instructions for Use 8.871.1003

10

Ensure that the instrument is never put into operation with fuses of a

different type as this could cause a fire!

For checking or changing fuses, proceed as follows:

1 Disconnect mains cable

Disconnect mains cable from mains connection plug 8.

2 Remove fuse holder

Take fuse holder 7 until it dangles (on the bottom of the fuse

holder is a small opening to grip it).

3 Take out fuse

Carefully take the fuse out of the fuse holder.

4 Insert fuse

Change fuses if necessary and reinsert in fuse holder.

5 Install fuse holder

Install fuse holder.

6 Mains switch

Figure 4: M

Mains ca

M

Th

6.2122

ains connection plug

2.2.2

.020 with plug SEV 12 (Switzerland,

)

6.2122.040 with plug CEE(7), VII (Germany,

)

6.2133.070 with plug NEMA 5-15 (USA,

)

which are three-cored and fitted with a plug with an earthing pin. If a dif-

ferent plug has to be fitted, the yellow/green lead (IEC standard) must

be connected to protective earth (protection class 1).

to switch instrument on and off:

I = ON 0 = OFF

7 Fuse holder

8 Mains connection plug

6

7

8

ble and mains connection

ains cable

e instrument is supplied with one of three mains cables