beautypg.com

2 parts and controls, Front view 853 co2 suppressor, Parts and controls – Metrohm 853 IC CO2 Suppressor User Manual

Page 8: Front view 853 co, Suppressor, Figure 1

background image

1 Introduction

853 CO

2

Suppressor / Instructions for Use 8.853.8001EN

2

1.2 Parts

and

controls

In this section you will find the numbers and designations of the parts

and controls of the 853 CO

2

Suppressor. The numbering applies

throughout the instructions for use, i.e. bold numbers in the text (e.g.

3

) refer to the parts and controls illustrated here.

1.2.1

Front view 853 CO

2

Suppressor

Figure 1: Front view 853 CO

2

Suppressor

1 Eluent flow outlet "Eluent out"

2 Eluent flow inlet "Eluent in"

3 LED "Power"

Lights up green when the pump is run-

ning.

4 LED "Status"

Lights up orange when the pump is

running and the pressure in the vacu-

um chamber is too high or too low.

After the pump has been switched on it

takes a few seconds for the vacuum to

be established.

5 LED "Ready"

Lights up green when the pump is run-

ning and the pressure in the vacuum

chamber is within the ideal working

range.

6 Aspiration opening "Air in"

Opening for drawing in CO

2

-free air

(through CO

2

Absorber Cartridge).

7 PEEK flow reduction capillary

6.1831.130
The length of the flow reduction capilla-

ry provides an optimal flow and should

not be altered.

8 Pressure screw

9 Coupling

6

7

8

9

1

2

5

4

3