5 appendix, 1 technical data, Appendix – Metrohm 853 IC CO2 Suppressor User Manual
Page 21: Technical data, Section 5 appendix
5.1 Technical data
853 CO
2
Suppressor / Instructions for Use 8.853.8001EN
15
5 Appendix
5.1 Technical
data
Provided that nothing to the contrary is mentioned, the
published data represents typical values for the 853 CO
2
Suppressor at an ambient temperature of 25°C.
Vacuum
Pressure range
< 65 mm Hg
Establishment time after start < 30 s
Flow range
Recommeded flow range 0.1…1.0
mL
Degassing capillary in 853 CO
2
Suppressor
Material
Teflon AF
Solvent resistance
Very good resistance to most solvents (excep-
tion: PFCs)
Ambient temperature
Nominal working range
+5...+45 °C (at max. 85% rel. humidity)
Storage
-40...+70 °C
Transport
-40...+70 °C
Power supply
Voltage
100...240 V (± 10%)
Frequency
50 / 60 Hz
Power consumption
15 W
Fuses
Overload protection
Safety specifications
Construction / Testing
EN/IEC/UL 61010-1,
CSA-C22.2 No. 61010-1,
Protection code IP20,
Protection class III
Safety information
These Instructions for Use contain information
and warnings that must be observed by the user
in order to ensure the safe operation of the in-
strument.