beautypg.com

3 instrument problems, Care and maintenance, 1 care – Metrohm 853 IC CO2 Suppressor User Manual

Page 19: Instrument problems, Care, 4 care and maintenance

background image

4.3 Instrument problems

853 CO

2

Suppressor / Instructions for Use 8.853.8001EN

13

4.3 Instrument

problems

Problem

Cause

Remedy

LED "Power" (1) does
not light up when in-

strument is switched
on.

Power supply not con-

nected


6.2152.020 Power sup-

ply faulty or electronics
fault in 853 CO

2

Suppres-

sor

• Connect the 853 CO

2

Suppres-

sor to the mains supply via the

external power supply (see
Section 2.2)

• Contact Metrohm Service.

Neither LED "Status"
(2) nor LED "Ready"
(3) lights up when in-

strument is switched on
(LED "Power" (

1) lights

up).

LEDs faulty

• Contact Metrohm Service.

Vacuum out of working
range. LED "Ready" (3)

does not light up after
instrument has been
switched on for at least
30 s.

Leak in system (vacuum

too weak)

H

2

O Absorber Cartridge

blocked (vacuum too

strong).

• Contact Metrohm Service.

• Regenerate the H

2

O Absorber

Cartridge (see Section 4.4.4

or 8.108.1046 Leaflet)

4.4 Care

and

maintenance

4.4.1 Care

The 853 CO

2

Suppressor requires adequate care. If it becomes excessi-

vely dirty then this could interfere with its functions and shorten the

working life of its robust mechanism and electronics.
Spilt chemicals and solvents must be wiped up immediately. The elec-

trical connections on the rear panel of the instrument (the mains con-

nection in particular) should be protected against contamination.

Although constructive measures to a large extent prevent liquid pene-

tration, if aggressive media should nevertheless penetrate the interior

of the instrument then the mains plug of the external 6.2152.020

Power supply should be pulled out immediately to prevent serious

damage to electronic components. If such damage should occur

please contact Metrohm Service.

The instrument should only be opened by specialists from Metrohm.