4 Troubleshooting - Problems

853 CO

2

Suppressor / Instructions for Use 8.853.8001EN

12

4 Troubleshooting - Problems

4.1

Remedying faults and problems

If difficulties occur during analyses with your IC system then it is best to

search for their causes in the following sequence: column pump

eluent IC system. In the Instructions for Use of your Modular IC

system or Compact IC you will find an overview of possible faults to-

gether with their causes and remedies.
In addition to these general problems, the following section covers

those problems which could arise from the use of the 853 CO

2

Supp-

ressor.

4.2 Chromatography

problems

Problem

Cause

Remedy

Poor retention time

reproducibility

Leak in the instrument.

Blockage in flow path.

• Contact Metrohm Service.

• Check the capillary connections

and replace any compressed
capillaries

Noisy or unstable
baseline

Leak in the instrument.

Blockage in flow path


• CO

2

cartridge exhausted

• Vacuum pump faulty

• Contact Metrohm Service.

• Check the capillary connections

and replace any compressed
capillaries

• Replace CO

2

cartridge (see Sec-

tion 4.4.3)

• Contact Metrohm Service.

High pressure in the
system

Pressure screws

tightened up too far.

• A liquid-transporting

component in the device
is blocked.

• Loosen pressure screws slightly

or replace connection (cut new

capillary end)

• Contact Metrohm Service.

Poor peak shape

Dead volume in system • Check capillaries