5 connection of capillaries, 6 cartridges, Connection of capillaries – Metrohm 853 IC CO2 Suppressor User Manual

Page 16: Cartridges

background image

2 Installation

853 CO

2

Suppressor / Instructions for Use 8.853.8001EN

10

2.5 Connection

of

capillaries

The 853 CO

2

Suppressor is connected between the suppressor module

«MSM II» and the detector block.

1 Suppressor module «MSM II» – 853 CO

2

Suppressor

• Connect the "Suppressor outlet capillary for eluent" (marked

with "Detector") to the "Eluent in" inlet (2) of the 853 CO

2

Suppressor using a 6.2744.010 Pressure screw.

2 853 CO

2

Suppressor – Detector block

• Connect the "Inlet capillary to detector block" to the "Eluent

out" outlet (1) of the 853 CO

2

Suppressor using a 6.2744.010

Pressure screw.

2.6 Cartridges

Effective CO

2

removal requires that the air pumped through the de-

gassing cell should contain as little CO

2

as possible. This is achieved

by drawing the air through a 6.2837.000 CO

2

Absorber Cartridge.

The CO

2

Absorber Cartridge can become blocked by moisture. This

can be prevented by connecting a 6.2837.010 H

2

O Absorber Car-

tridge in front of it.

The cartridges are installed as follows:

1 Remove caps

• Remove the sealing caps at the inlet and outlet of each

cartridge.

2 Connect CO

2

Absorber Cartridge

• Insert the CO

2

Absorber Cartridge (3-layer filling, blue-brown-

gray) in the coupling 9.

3 Connect H

2

O Cartridge – CO

2

Cartridge

• Insert the 6.1808.190 Adapter in the CO

2

Absorber Cartridge.

• Attach the 6.1801.140 PVC tubing to the 6.1808.190 Adapter.
• Insert the H

2

O Absorber Cartridge (filled with orange drying

beads) into the 6.1801.140 PVC tubing.

4 Attach the cartridges

• Attach the two cartridges to the left-hand mounting support of

the 861 Advanced Compact IC or 820 IC Separation Center

using two 6.2027.070 Holders.