2.4 Connection to 830 IC Interface
853 CO
2
Suppressor / Instructions for Use 8.853.8001EN
9
2.4
Connection to 830 IC Interface
Connect an Event line of the 830 IC Interface to the remote interface 12
of the 853 CO
2
Suppressor using a cable 6.2128.180.
Figure 4: Connection between the 853 CO
2
Suppressor and 830
IC Interface
In the PC program «IC Net» the 853 CO
2
Suppressor does not have its
own device driver. It can be controlled via the 830 IC Interface. This
means that the 830 IC Interface must be included in your system and
configured for controlling the 853 CO
2
Suppressor:
1 Open «IC Net» system
• In the PC program «IC Net» open the system in which the 830
IC Interface is to be included.
2 Include the 830 IC Interface in the system
• Use Setup/New devices/Link to existing device to select
the 830 IC Interface (the 830 IC Interface must already have
previously been added to the "Workplace", see IC Net
Instructions for Use, Section 5). The device symbol for the 830
IC Interfaces is added to the system.
3 Open the "Events setup" tab
• Double-click on the device symbol of the 830 IC Interfaces to
open its configuration menu and select the Events setup tab.
4 Set the control line for the 853 CO
2
Suppressor
• The value for the Event line used to control the 853 CO
2
Suppressor must be set to 1. In the example shown above this
is Event Line 1, see Figure 4.
The 853 CO
2
Suppressor has now been included in your system and
can be controlled by the PC program «IC Net». It is started together
with the other system hardware with System/Control/Startup hard-
ware and closed down with System/Control/Shutdown hardware.