6.1 Problems and their solutions
■■■■■■■■■■■■■■■■■■■■■■
80
■■■■■■■■
850 Professional IC Cation – LP Gradient
Problem
Cause
Remedy
Sample – Gas bubbles in
the sample.
Use the sample degasser (see Chapter 3.14,
page 44) (if available).
Very noisy baseline
High pressure pump – con-
taminated pump valves.
Clean pump valves (see "Cleaning the inlet
valve and outlet valve", page 67).
Eluent – Leak in the eluent
path.
Check the eluent path
Eluent – Blockage in the
eluent path.
Check the eluent path
High pressure pump –
defective piston seals.
Replace piston seals (see "Replacing the piston
seal", page 64).
Pulsation damper not con-
nected.
Connect the pulsation damper (see Chapter
3.13, Page 43).
Pulsation damper not con-
nected.
Connect pulsation damper (see Chapter 3.13,
page 43).
Background conduc-
tivity too high
Incorrect eluent.
Change eluent (see Chapter 5.4.2.3, page 61).
Data of the separa-
tion column cannot
be read.
Column chip contamina-
ted.
Clean contact surfaces of the column chip
(with alcohol).
Column chip defective.
1. Save column configuration in MagIC Net.
2. Notify Metrohm Service.
Individual peaks
greater than expec-
ted
Sample – Carry-over of the
samples from previous
measurement.
Rinse system longer between two samples.
Poor reproducibility
of the retention
times
Eluent – Leak in the eluent
path.
Check the eluent path
Eluent – Blockage in the
eluent path.
Check the eluent path
Chromatograms
have poor resolution
Separation column –
diminished separating effi-
ciency.
■
Regenerate separation column (see Chap-
ter 5.11.4, page 77).
■
Replace separation column (see "Connect-
ing and rinsing the separation column",
page 54).