Page 50: Mixing coil for low pressure gradients connection

background image

3.11 High pressure pump

■■■■■■■■■■■■■■■■■■■■■■

40

■■■■■■■■

850 Professional IC Cation – LP Gradient

1

2

3

Figure 21

Mixing coil for low pressure gradients connection

1

Inlet valve holder

2

PEEK pressure screw, short 6.2744.070

3

Mixing coil for low pressure gradients
6.2758.020

3.11.3

Deaerating the high pressure pump

The high pressure pump will only operate perfectly if the pump head con-
tains no more air bubbles. Therefore it must be deaerated during initial
start-up and after every change of eluent.

Caution

The high pressure pump must not be deaerated before the initial start-
up .

Deaerate the high pressure pump as follows (see Figure 22, page 41):

Deaerating the high pressure pump

The instrument must be connected to the PC and switched on to deaerate
the high pressure pump.