825 Lab Link, Instructions for Use
7
2.3 Connection of the 825 Lab Link
Connect the 825 Lab Link to your switched-off Titrando system as
shown in Section 1.1 and connect it to the network:
• To do this connect the USB RS 232 Box to a USB connection
of the Titrando by using the 6.2151.030 USB Cable.
• Connect one RS 232 connection of the USB RS 232 Box to RS
232 connection 1 of the 825 Lab Link by using the 6.2134.040
Cable.
• Connect the 825 Lab Link to the local network by using an
RJ45 Ethernet cable at connection 3.
• For MSB connection 7 of the 825 Lab Link you can use any
free MSB connection of the Titrando or a peripheral device.
This connection is then made available again on the 825 Lab
Link to exactly the same extent (MSB connection 6).
2.4 Configuration of the Touch Control
The Touch Control used must be set for data transmission to an
external system:
1. Switch the Touch Control on.
2. If no PC/LIMS has been set up under System/Device
manager, then insert it with [New].
3. Select PC/LIMS and operate [Edit]. Set "Ext. System
access" as the communication mode.
4. You should also set the RS 232 interface according to
the selected connection at the USB RS 232 Box.
5. The Touch Control must then be switched off in order to store
these settings.