Figure 34, Install pump tubing connection without filter – Metrohm 850 Professional IC Anion MCS LP-Gradient User Manual
Page 71
■■■■■■■■■■■■■■■■■■■■■■
3 Installation
850 Professional IC – Anion – MCS – LP Gradient
■■■■■■■■
61
1
2
3
Figure 34
Install pump tubing connection without filter
1
Union nut
2
Adapter
3
Tubing olive
■
Slide union nut (34-1) onto the pump tubing.
■
Select a suitable adapter (34-2) and slide it onto the pump tubing.
The type of adapter depends on the pump tubing (see Table 2,
page 61).
■
Place the tubing olive (34-3) onto the pump tubing.
■
Screw the union nut (34-1) onto the tubing olive (34-3).
4 Inserting the pump tubing
■
Press the contact pressure lever all the way down.
■
Place the pump tubing in the tubing cartridge. The stoppers
(32-3) must snap into the corresponding holders of the tubing
cartridge.
5 Inserting the tubing cartridge
■
Hang the tubing cartridge in the mounting pin and press in the
cartridge holder until the snap-action lever snaps in.
6 Connecting the capillaries
■
Screw the respective capillaries tightly to the two tubing olives
with PEEK pressure screws (32-1).
Table 2
Pump tubings and suitable adapters
Pump tubing
Adapter
6.1826.020 (blue/blue)
6.1826.310 (orange/green)
6.1826.320 (orange/yellow)
6.1826.330 (orange/white)