Figure 33, Install pump tubing connection with filter – Metrohm 850 Professional IC Anion MCS LP-Gradient User Manual
Page 70
3.19 Peristaltic pump
■■■■■■■■■■■■■■■■■■■■■■
60
■■■■■■■■
850 Professional IC – Anion – MCS – LP Gradient
3 Connecting the pressure side
Note
Depending on the use of the peristaltic pump, on the pressure side
you can either connect:
■
Case A: a 6.2744.180 pump tubing connection with filter
(see Figure 33, page 60) or
■
Case B: a 6.2744.160 pump tubing connection without filter
(see Figure 34, page 61).
For pumping the auxiliary solutions to the MSM or to the SPM, a
6.2744.180 pump tubing connection with filter must be used.
Case A: 6.2744.180 pump tubing connection with filter:
1
2
3
Figure 33
Install pump tubing connection with filter
1
Union nut
2
Adapter
3
Tubing olive with filter holder
■
Slide union nut (33-1) onto the pump tubing.
■
Select a suitable adapter (33-2) and slide it onto the pump tubing.
The type of adapter depends on the pump tubing (see Table 2,
page 61).
■
Place the tubing olive with filter holder (33-3) onto the pump tub-
ing.
■
Screw the union nut (33-1) onto the tubing olive (33-3).
or
Case B: 6.2744.160 pump tubing connection without filter: