Figure 50, Msm – components – Metrohm 850 Professional IC Anion MCS LP-Gradient User Manual
Page 108

5.11 Metrohm Suppressor Module (MSM)
■■■■■■■■■■■■■■■■■■■■■■
98
■■■■■■■■
850 Professional IC – Anion – MCS – LP Gradient
1
2
3
4
5
Figure 50
MSM – Components
1
Union nut
2
MSM connecting piece 6.2832.010
3
MSM rotor A 6.2832.000
4
MSM housing
5
Slot in the MSM housing
Cleaning the MSM
Clean the MSM as follows:
1 Disconnecting the MSM from the IC system
■
Switch off the instrument.
■
Disconnect the MSM from the separation column, peristaltic
pump and detector.
2 Dismantling the MSM
■
Unscrew union nut (50-1) from the MSM housing (50-4).
■
Pull MSM connecting piece (50-2) and MSM rotor A (50-3) out of
the MSM housing (50-4). The MSM connecting piece and MSM
rotor A normally stick to one another - if this is not the case: Take
a sharp object, insert into the slot (50-5) in the MSM housing,
and pull out MSM rotor A (50-3) in this way.
■
Detach the MSM connecting piece (50-2) from MSM rotor A
(50-3).
3 Cleaning the supply and discharge lines
■
Connect in turn each of the 6 capillary tubings fastened on the
MSM connecting piece (50-2) on the high pressure pump (see
Chapter 3.12, page 42) and pump through ultra pure water.