beautypg.com

2 optional accessories, Optional accessories – Metrohm 850 Professional IC Anion MCS LP-Gradient User Manual

Page 146

background image

9.2 Optional accessories

■■■■■■■■■■■■■■■■■■■■■■

136

■■■■■■■■

850 Professional IC – Anion – MCS – LP Gradient

Qty.

Order no.

Description

1

8.850.8053EN Manual 850 Professional IC, 2.850.2210 -

Anion - MCS - LP Gradient, English

9.2

Optional accessories

2.850.2210 850 Professional IC –Anion – MCS – LP Gradient

Order no.

Description

6.1014.000

Metrosep A Trap 1 column

Trap column used for anion determinations with gradient

Housing material:

PEEK

Base:

Polystyrol/Divinylbenzol-Copolymer with qua-
ternary ammonium groups

Particle size (µm):

570 µm

Organic modifier:

0 - 20 %

Maximum pressure:

250 MPa

pH range:

1 - 14

Column dimensions:

100 x 4.0

Regeneration:

Rinsing with 30 mL 0.5 mol/L Na

2

CO

3

6.1826.310

Pump tubing LFL (orange/green) 3-stopper

Pump tubing for bromate determination using the triiodide method.