beautypg.com

Initial setup, Figure 15 calibration potentiometer – Bird Technologies 4391A User Manual

Page 31

background image

19

Initial Setup

Note: On a clean flat workspace perform the following setup proce-
dure:

1.

Remove the front panel. Refer to "Front Panel Removal" on page 23.

2.

Insert a DC feed-in adapter into the FWD element of the socket.

3.

Insert a metal dust plug (supplied with the instrument) into the RFL element
socket.

4.

Connect a BNC to BNC cable assembly from the output of calibrator/source
to the DC feed-in adapter.

5.

Set the RF Power Analyst range switches to the 10 and x100 positions.

6.

Connect AC power and place the LINE/OFF/BATT switch to the LINE setting.

7.

Switch on the Calibrator/Source.

8.

Allow it to warm up for at least one hour.

9.

Set the Calibrator/Source to output 30 ±0.05 microamps.

10. Adjust potentiometer, R25 (Figure 15 on page 19), until the RF Power Ana-

lyst Displays 1000 ±5.

Figure 15 Calibration Potentiometer