Caution statements – Bird Technologies 8650B Series User Manual
Page 5
iii
Caution Statements
The following equipment cautions appear in the text whenever the equipment is
in danger of damage and are repeated here for emphasis.
CAUTION
Before any RF operation of the load is attempted, the transmitter interlock and
AC line attachment to the equipment must be made. First attach the
transmitter interlock connections to the two binding posts on the front panel
and then connect the power cable.
CAUTION
The 115 VAC or 230 VAC line may be connected, but the AC power must not be
turned on unless coolant is in the system and filled to the correct level. Damage
to the pump will result otherwise.
CAUTION
Use only distilled water or the supplied ethylene glycol as coolant. Do not use
tap water, automotive antifreeze, sealants, or leak stopping material. Use of
these materials will damage the instrument and void all warranties.
CAUTION
Operation without sufficient coolant can damage the unit.
CAUTION
Do not block air flow. Air enters through the side openings of the Moduload and
exhausts through the top.
CAUTION
Do not operate the load without proper coolant levels. Operation without
sufficient coolant can damage the unit.
CAUTION
Do not operate the load without connecting the interlock. Even momentary application
of power to the load while the cooling circulation is off, or possibly functioning
improperly, will cause almost immediate destruction of the resistor element.