Bird Multi-Sensor Cal Cart System

8

Error Codes

The Bird 4421 displays error codes when the RF power is either below the
selected range (underrange) or above the selected range (overrange). Figure 5
displays the error codes and Figure 6 lists the function limits.

Figure 5 Error Codes

Figure 6 Function Limits

Audible Warning

If the RF power level exceeds 120% of the power sensor’s maximum power
capability, the power meter will sound a warning buzzer.

Symbol

Explanation

Value greater than overrange limit of function

Value less than underrange limit of function

Function

Limit

Error

FWD, RFL

Power > 199.9% of full scale or 120% of top
range

Overrange

FWD dBm, RFL dBm

Power > 120% of full scale
Power < 3% of low range

Overrange
Underrange

SWR

FWD < 20% of low range
FWD – RFL = 0

Underrange
Overrange

Return Loss

FWD < 20% of low range
RFL < 20% of low range
Return Loss > 40 dB

Underrange
Underrange
Underrange