Chapter 2 Time Recorder Extension Module
6.
Select a parameter value and press the Left arrow.
7.
Press F3 (Save) to save the current settings to a file.
Save a Time Recorder Setup and Measurement on page 23
Use the descriptions in
to help you configure the Time Recorder
parameters.
Table 2 - Time Recorder Parameter Descriptions
Parameter
Description
Values/Comments
Input
The input channels for the recorder measurement.
This parameter affects the Freq Range maximum frequency setting and
the recording’s maximum recording time.
• X - 40 KHz maximum Freq Range
• X & Y - 20 KHz maximum Freq Range for each channel
• X & Tacho - 20 KHz maximum Freq Range for each channel
X (default)
X and Tacho
X & Y
X & Y and Tacho
X, Y, & Z
X, Y, Z & R
X Type
Sets the vibration measurement type used on Channel X.
Your selection determines the available options and engineering units for
subsequent set-up parameters.
Accel (g) (default)
Accel (m/s
2
)
Vel (mm/s)
Vel (ips)
Disp (mils)
Disp (µm)
Volts (V)
Volts AC
Press. (Pa)
Force (N)
Force (lBf)
Sens1
The sensitivity of the transducer in millivolts (mv) per Engineering Units
for channel X.
The sensitivity value is included with the transducer’s documentation or it
may be imprinted on the side of the transducer.
mV/EU