beautypg.com

Specification table – Yamato Scientific BH500 Thermo-Elite User Manual

Page 32

background image

29

Specification table

BH400

BH500

Range of temperature

setting (℃)

-30.0∼+120

-30.0∼+220

Range of operation

temperature (℃)

Ambient temperature +10∼100

Ambient temperature +10∼200

Temperature stability (℃)

±0.01℃∼0.02℃ (Water,at ambient temperature +15∼80℃, RT25℃, No-load)

±0.05℃∼0.1℃ (silicon oil , KF96 / viscosity:50cs , liquid temperature at 70∼200℃)

Pump capacity

Max. pump head : 3(mH

2

O), Max. stream flow : 20 (

"

/min)

(Refrigerant : Water , Water temperature : 25℃)

Temperature adjustment

machine

IV-CR type controller (Zero crossover switching, PID microprocessor control)

Sensor

Pt100Ω class A (three line type)

Temperature set method

Digital setting depending on ▲▼keys

Temperature display method

Digital on a main display (A set temperature is displayed to a sub display)

Resolution of setting and display

temperature

0.1℃

Other displays

Drive monitor (represents the state of control)

Timer function

1 minute to 99 hours 59 minutes , 100 hours to 999 hours

Timer resolution

1 minute or 1 hour

Drive function

Fixed temperature drive, Auto start drive, Auto stop drive, Program drive : Arbitrary

pattern, Max.16 segment (repetition and ramp operation, etc.)

Output terminal

(7 pin connector)

Other incidental functions

Memory backup

Key lock

Multiplication

time function

( up to 49999

hours)

Holding

Function that

enable to change

the strength of jet

stream (10 step

changeability)

Calendar timer

Clock

Analogue

External

warning output

Time-up output

External

communication

(RS422A

correspondence)

Communication

terminal

(8pin connectors)

Heater (made of SUS)

1.0kw

1.2kw

stirring mechanism

Stirring by jet stream

Other compositions

Drain (with plug) , Stainless flexibility tube with heat insulation ,

Inlet for Immersion cooler (BE series) , cooling coil (O.D φ9.5mm)

Earth leakage breaker

Thermal protector (for stirring motor)

The float type High/Low liquid level detector (Low level only for BH400)

Self-diagnosis function (Automatic overheat protector, The sensor is abnormal,

Heater disconnection, Triac short-circuit, Main relay is defective, Buzzer warning

in abnormal circumstances, The refrigerator is abnormal)

Safety device

none

independent temperature overheating

prevention device (BH500)

Capacity of reservoir (

"

)

13

Size of reservoir

(W

×D×H mm)

389Ч226Ч150

The outside size

(W

×D×H mm)

654Ч492Ч615

Power supply

AC100V 11A

AC100V 13A

Standard attached goods

Stainless hose entrance for external circulation (Rc1/2×Outside diameterφ14mm)×2

Digital printer

independent temperature overheating prevention device (for

BH400)

Attached goods outside standard

By-pass tube for external circulation, Insulated hose for external circulation

Cable for external communication, Personal computer connection interface,

(RC23), Multi wick cable for output (All model commonness)