beautypg.com

Safety device and error code – Yamato Scientific BH500 Thermo-Elite User Manual

Page 28

background image

25

Safety device and error code

The self-diagnosis function built into by the controller and the independent safety device from

the controller are installed in this machine. The table below shows the cause and the

processing method when the safety device operates. When the safety device operates, the error

code is displayed to a main display. Deal with it according to the processing method shown

here.

Safety device

Display

Cause and counter-medsures

Notes

Earth leakage

breaker with

over current

protector

No display

Report to our service office and check
the cause of problem.

Motor thermal

protector

No display

Overheating of motor:
Report to our service office.

Automatic

overheat

protector.

No display

Heating of samples:
Reduce the amount of the sample.

Sensor

malfunction

detector

TROUBLE lamp flashes

flashes

Break in temperature sensor circuit:
Report to our service office.

Triac short-

circuit detection

TROUBLE lamp flashes

flashes

Short-circuit of Triac:
Report to our service office.

Heater short-

circuit detection

TROUBLE lamp flashes

flashes

Heater circuits disconnected:
Report to our service office.

Incorrect setting of independent
temperature overheating prevention
device:
Set correctly.

Heating of samples:
Reduce the amount of the sample.

Independent

temperature

overheating

prevention

device

TROUBLE lamp flashes

flashes

Malfunction of the independent
temperature overheating prevention
device:
Report to our service office.

BH500 and
the machine
equipped with
independent
temperature
overheating
prevention
device

Main relay

defect detection

TROUBLE lamp flashes

flashes

Malfunction of main relay
Report to our service office.

Decrease in water level:
Supply water.

Rise of water level (thermal expansion):
Decrease liquid from reservoir.

Water level

abnormality

detection

TROUBLE lamp flashes

flashes

Malfunction of water level detector:
Report to our service office.

The unit
resume to
working when
liquid level
becomes
normal.

This manual is related to the following products: