ColonyDoc-It Imaging Station

22

The new window will display the

Total colonies

counted.

Fluorescent Colony Counting

Ensure that the doors are attached to the

ColonyDoc-It or turn off the surrounding lights.

Turn the

Epi Lighting switch on the ColonyDoc-It

to

Blue.

Insert the

Green Fluorescent Protein (GFP)

emission filter into either tray (refer to the
Installing the Filter

section of this manual for

instructions) and ensure that the filter is positioned

below the camera. (If the GFP filter is in filter

position one, the filter slider should be moved all

the way to the right position.)

Insert the fluorescently stained sample into the

ColonyDoc-It.

Select the appropriate

Plate Diameter, the

Smallest Colony information and choose the

Fluorescent lighting function of Colony Lighting

in the Doc-It Colony Counter software.

The slider bar may be adjusted from

Dim to

Bright to provide optimum results.

Select

Preview to position the plate inside the red

circle guides and then select

Capture when the

image is positioned properly.