Counting colonies, Quick capture – UVP ColonyDoc-It Imaging Station User Manual
Page 13

ColonyDoc-It Imaging Station
13
Counting Colonies
Quick Capture
Place the Petri dish to be counted on the transillumination surface.
Select the appropriate lighting option for the sample. If unsure, turn the selector switch (located on the front of the
colony counter) to
Base Lighting > Trans and remove the black plate (if present). The Trans lighting option will
apply to most applications.
If the colony counter is not already open, double click onto the
Doc-It Colony Counter icon on the desktop.
Ensure that the
Camera Control module is active. See the image of the screen below noted by the green circle.
The text must be black as shown. If the
Camera Control is inactive (with grey text) proceed to the troubleshooting
section: Inactive Camera Window.
Select the Petri dish plate diameter from the
Camera Control module. The choices are 6cm, 10cm, and 15cm.
Select the
smallest colony size for this plate. The choices are <1mm and 1mm+. If <1mm is selected the
smallest colony size is below 1mm. The count will be faster if 1mm+ is chosen.
Choose the
White Light option from Colony Lighting. The Dim to Bright slider will be disabled unless using the
Fluorescent Colony Lighting option. If counting fluorescent colonies go to the Fluorescent Colony Counting
section of the manual.