Service Procedures

Return Procedure

A Returned Goods Authorization (RGA) number must be obtained from UVP Customer Service before
returning any product.

Replacement Parts and Accessories

To order accessories (including additional emission filters) or replacement parts for the GelDoc-It System,
contact UVP’s offices.

Part Description

Part Number


UV Handheld Lamps:

UV Handheld Lamp, 254/365nm, 4 watt (115V)

95-0021-12 (2 Recommended)

UV Handheld Lamp, 254/365nm, 4 watt (230V)

95-0021-10 (2 Recommended)


Fuses:

Fuse, 3.15A (for Darkroom)

56-0022-04 (2 Required)

Fuse, 2A (for Transilluminators)

56-0002-01 (2 Required)

Fuse, 1A (for GelCam 310 Motorized Lens Controller Box)

56-0022-02 (2 Required)


Emission Filters:

Filter, Ethidium Bromide, 50mm Square

38-0220-01

Filter, SYBR Green, 50mm Square

38-0219-01

Filter, SYBR Gold, 50mm Square

38-0221-01


Transillumination Accessories:

LED White Light Plate

95-0476-01

White Light Converter Plate, 21x26cm

38-0191-01

Visi-Blue Converter Plate, 21x26cm

38-0200-01


Gel Accessories:

Gel-Cutter

85-0002-01

Gel-Ruler

85-0003-01

Gel-Scooper

85-0006-01

Gel-Tray, small

85-0007-01

Gel-Sentry DNA Preparation Plate

97-0076-01

Fluorescent Standard Step Tablet

33-0014-02


Protective Equipment:

Spectacles, UV Blocking (UVC-303)

98-0002-01

Goggles, UV Blocking (UVC-503)

98-0002-02

Faceshield, UV Blocking (UVC-803)

98-0002-04

Troubleshooting

No Power to the Darkroom or Transilluminator

1. Recheck the main power cord connection to the GelDoc-It

2

/ChemiDoc-It

2

darkroom as well as the

power cables between the darkroom and transilluminator, LED White Light Plate or optional UV
handheld lamps.

2. Check fuses, located at the back of the unit near the power port. A flat-head screwdriver will be

required. Turn the fuseholder cap counterclockwise and the fuse holder will pop out. Inspect the
thin wire within the glass fuse to see if there is a break in the wire. If so, replace the fuse(s). If fuses
are blowing repeatedly, contact UVP Technical Support for additional troubleshooting.