UVP ChemiDoc-It 2 Imagers User Manual
Page 11
GelDoc-It
2
and ChemiDoc-It
2
Imagers
11
7. Plug the serial cable from the motorized lens
into the lens control box.
8. Connect the USB cable to the PC connection
port on the lens control box. Leave the other
end of the USB cable unplugged until
installation of VisionWorksLS is complete.
9. Connect the power supply to the lens control
box and into the power outlet.
MegaCam 810, OptiChemi 610 and BioChemi 510 Cameras
Specifications
MegaCam 810
OptiChemi 610
BioChemi 510
Type
CCD, Monochrome
CCD, Monochrome
CCD, Monochrome
Resolution
3296 x 2472
2184 x 1472
2.1MP
Bit Depth
16-bit
16-bit
16-bit
Quantum Efficiency
50% peak
86% peak
50% peak
Cooling
-
35˚C from Ambient
-
50˚C from Ambient
-35
˚C from Ambient
Binning
Yes, 1x1 thru 8x8
Yes, 1x1 thru 10x10
Yes, 1x1 thru 8x8
PC Interface
USB
USB
USB
Components
Note: Some components may not be shipped with all systems.
Note: The MegaCam 810, OptiChemi 610 and BioChemi 510 cameras use many of the same
components and are installed in a similar manner. Therefore, the installation instructions herein will
refer to all three cameras interchangeably.
PC Connection on the
Lens Control Box
Power supply on the
Lens Control Box