Ii. site preparation – SciGene Little Dipper Processor for FISH User Manual
Page 6
Little Dipper
®
Processor for FISH— USER MANUAL
March 2013
LD for FISH manual 2.0; software 1.4 for FISH
5
II. SITE PREPARATION
The Little Dipper Processor for FISH requires a level surface area of at least 30 x 22 inches
(76
x
56
cm) with a vertical clearance of 30 inches (76 cm) (Fig. 1). Locate the unit within 3
feet (1
meter) of a single power outlet that delivers at least 1700 Watts/20 Amps of
constant power.
Clear the benchtop before unpacking any cartons. Allow unimpeded access to the
Emergency Off (EMO) and main power switches on the right side of the instrument.
Figure 1. Allow proper clearance for the instrument.