beautypg.com

RKI Instruments GX-2009 User Manual

Page 25

background image

Read Commands • 25

5. The program will retrieve the instrument’s parameter

configuration and most recent calibration information then
add them to an existing .csv file or save a new .csv file with
the instrument’s parameter configuration and most recent
calibration information as described in step 3 above.

6. Use a spreadsheet program such as Microsoft Excel to open

and view the saved .csv files.

7. The instrument’s parameter configuration is now loaded in

the User Setup Program and the parameters are available
for updating in the GX-2009 Status area of the Main Window
and in the Factory Settings Window. The Factory Settings
Window is accessed using the Detail Settings command
button in Setting Commands. See “Detail Settings” on
page 44 for a description of Detail Settings.

8. See step 5 through step 9 in “Instrument Information” on

page 20 for instructions to update parameters in the GX-
2009 Status area of the Main Window.